GoodKnight® 420G Patient Manual Ref. : M-139DFU02-10 Revision C
Revision A 4 GoodKnight 420G - Patient Manual DESCRIPTION OF THE DEVICE The GoodKnight 420G is powered from an external power supply and equippe
Revision A GoodKnight 420G - Patient manual 5 The items in the Figures below are described on page 4. Figure 2 – Top view Figure 3 – Genera
6 GoodKnight 420G - Patient Manual This page is left blank intentionally
Revision A GoodKnight 420G - Patient manual 7 INSTALLATION This chapter shows you how to connect the various accessories and cables to your device.
Revision A 8 GoodKnight 420G - Patient Manual Fitting an oxygen adaptor Substitute the following step for step 1 of the standard installation proce
Revision B GoodKnight 420G - Patient Manual 9 Installing a humidifier Please refer to the humidifier user instructions for your for further informa
Revision B 10 GoodKnight 420G - Patient Manual Installing an external battery The GoodKnight 420G can be powered by a 12-Volt car battery by using
Revision A GoodKnight 420G - Patient Manual 11 USE This chapter shows you how to use the device after installing it as described in the previous ch
Revision A 12 GoodKnight 420G - Manuel patient WARNINGS • When the device is in operation mode, check that it delivers an airflow. If no airflow i
Revision A GoodKnight 420G - Patient Manual 13 You have to press this button the number of times required for displaying the screen below: 4.0cmH2O
Revision C ii GoodKnight 420G - Patient Manual List of Patient Manual revisions GoodKnight® 420G The pages below are included in the American
Revision A 14 GoodKnight 420G - Manuel patient Accessing information on the device The display enables you to view the different information on the
Revision A GoodKnight 420G - Patient Manual 15 In this case, the display again shows one of the following two screens indicating the status of your
16 GoodKnight 420G - Manuel patient This page is left blank intentionally
Revision A GoodKnight 420G - Patient Manual 17 MAINTENANCE AND CLEANING INSTRUCTIONS In order to ensure that you use your GoodKnight 420G, safely a
Revision A 18 GoodKnight 420G - Patient Manual Nasal mask Please refer to the instructions for use that were sent to you with your nasal mask. Hum
Revision A GoodKnight 420G - Patient Manual 19 TROUBLE-SHOOTING PROBLEM PROBABLE CAUSE SUGGESTION Discomfort due to a feeling of the pressure b
Revision A 20 GoodKnight 420G - Patient Manual PROBLEM PROBABLE CAUSE SUGGESTION Recurrence of the first symptoms of sleep apnea syndrome. The
Revision B GoodKnight 420G - Patient Manual 21 TECHNICAL CHARACTERISTICS Device performance Pressure range of the GoodKnight 420G: 4 cmH2O to 20 cm
Revision C 22 GoodKnight 420G - Patient Manual Electrical characteristics of the power supply module Input voltage: 100 – 240 VAC (-10%, +10%), 50
Revision C GoodKnight 420G - Patient Manual iii The Puritain Bennett GoodKnight® 420G device (GK420G) is intended for treating patients with Sl
Corporate Headquarters: Nellcor Puritan Bennett Incorporated 4280 Hacienda Dr. Pleasanton, CA 94588-8604 Tel. +1 (925) 463-4000 Fax : +1 (925) 463-
iv GoodKnight 420G - Patient Manual This page is left blank intentionally
Revision A GoodKnight 420G - Patient Manual v TABLE OF CONTENTS BEFORE STARTING...
GoodKnight 420G - Patient Manual vi This page is left blank intentionally
Revision A GoodKnight 420G - Patient Manual 1 BEFORE STARTING Safety instructions In order to ensure that you use your GoodKnight 420G device safe
Revision A 2 GoodKnight 420G - Patient Manual Indications for use The GoodKnight 420G device is indicated for use in the treatment of obstructive
Revision A GoodKnight 420G - Patient manual 3 Checking the components In order to be able to use the GoodKnight 420G with its specific electrical p